330-1G LL Professional Combustion Analyzers Kits
Ordering Information
330-1 G LL Kit 1 Residential Analyzer:
analyzer, probe, power supply, spare filters and case….0563 3371 70
330-1G LL Kit 2 Residential Analyzer:
analyzer, probe, power supply, spare filters, case, IR printer and printer paper….0563 3371 75
330-1G LL Kit 3 Residential / Light Commercial Deluxe:
analzyer, probe, power supply, spare filters, case, IR printer, printer paper, smoke pump
test kit…500563 3303
330-1G LL Kit 4 NOx Kit:
analyzer, probe, power supply, spare filters, case, IR printer, printer paper, NO upgrade module (0 to 3000 ppm)…400563 3304
330-1G LL Kit 5 Residential / Commercial plus software w/ memory:
analyzer, probe, power supply, spare filters, case, IR printer, printer paper, USB cable EasyHeat software…400563 3305
Thanks to the new instrument functions, the Testoflue gas analyzers of the new testo 330-LL series now offer you even moreprofessional and reliable support. These are the main features: å¥High-resolution colour display for the graphic representation of yourmeasurement data å¥ Extended measurement menus åÐ such as gas pipe testing åÐfor a comprehensive analysis of your heating system å¥ Logger function for easylong-term recording of the measurement curve
- å¥ Graphic processing of measurement data
å¥ Instrument diagnosis function uses easy åãtraffic lightåÕåÕpresentation to enable comprehensive instrument diagnosis
å¥ Pressure measurement up to 300 mbar
å¥ User-defined fuels
å¥ TÌÏV-tested according to 1. BImSchV / EN 50379 Part 2 for O2, °C, hPaand CO with H2-compensation
å¥ Powerful memory management: 500,000 readings
å¥ IrDa/ Bluetooth interface for data transfer to Pocket-PC / laptop /printer
å¥ USB interface for data readout to a PC software
å¥ ZIV (Central Guild Association) driver for all standard industrysoftware packages
Included in delivery:
å¥ Option: Fine draught measurement, rResolution 0.1 Pa, measurementrange to 100 Pa (instead of the standard draught measurement)
å¥ Option fine differential pressure measurement
å¥ Option: NO sensor, meas. range 0 to 3000 ppm, 1 ppm resolution
å¥ Option H2-compensated CO cell
å¥ Option COlow sensor
å¥ Option Bluetooth
TechnicalData
| 500.000 values |
| Meas. Range | 0 to 500 ppm |
| Accuracy | ±5 ppm (0 to 100 ppm) ±5% of rdg. (>100 ppm) |
| Resolution | 1 ppm |
| Response time | Approx. 35 sec. |
| Probe type CO2 | |
|---|---|
| Meas. Range | 0 to 1 Vol. % 0 to 10.000 ppm |
| Accuracy | ±(50 ppm ±2% of rdg.) (0 to 5000 ppm) |
| Response time | Approx. 35 sec. |
| Probe type Draught measurement | |
| Meas. Range | -9.99 to +40 inH2O |
| Accuracy | ±0.02 inH2O or ±5% of rdg. (-0.50 to +0.60 inH2O) ±0.03 inH2O (+0.61 to +3.00 inH2O) ±1.5% of rdg. (+3.01 to +40.00 inH2O) |
| Probe type Exhaust gas loss measurement | |
| Meas. Range | 0 to 99.9% |
| Resolution | 0.1% |
| Probe type Electrochemical measurement CO | |
| Meas. Range | 0 to 4000 ppm |
| Accuracy | ±20 ppm (0 to 400 ppm) ±5% of rdg. (401 to 1000 ppm) ±10% of rdg. (1001 to 4000 ppm) |
| Resolution | 1 ppm |
| Probe type Electrochemical measurement NO | |
| Meas. Range | 0 to 3000 ppm |
| Accuracy | ±5 ppm (0 to 100 ppm) ±5% of rdg. (101 to 2000 ppm) ±10% of rdg. (2001 to 3000 ppm) |
| Resolution | 1 ppm |
| Probe type Electrochemical measurement O2 | |
| Meas. Range | 0 to 21 Vol. % |
| Accuracy | ±0.2 Vol. % |
| Resolution | 0.1 Vol. % |
| Probe type Efficiency Measurement (Eta) | |
| Meas. Range | 0 to 120% |
| Resolution | 0.1% |
| Probe type CO2-determination | |
| Meas. Range | 0 to CO2 max |
| Accuracy | ±0.2 Vol. % |
| Reaction time t90 | <40 sec. |